Rotary vane vacuum pump troubleshooting methods explain

In operation, rotary vane vacuum pumps often fail if operated improperly, for example, the vacuum of rotary vane vacuum pumps is reduced, smoke, fuel injection, oil leakage at oil seals, noise and the like, and the vacuum pump oil emulsification inside the pump . In case of such a situation, first of all do not panic, so after the shutdown according to the fault one by one excluded. The following is a rotary vane vacuum pump failure, I hope it is useful!

Rotary vane vacuum pump troubleshooting methods explain

First, rotary vane vacuum pump smoke and fuel injection: refers to the rotary vane vacuum pump in operation exhaust or fuel injection.

1, smoke, if the pump has just begun to run a smoke phenomenon, is normal, if the smoke is not normal for a long time. The solution: smoke shows the pump inlet, including pipes, valves, containers are repaired. Leakage after treatment, the smoke will end.

2, fuel injection, indicating a large number of leaks outside the air inlet, or even the air inlet exposure to the atmosphere. Solution: seal the pump inlet to pump operation, if you do not fuel injection, indicating a leak; exhaust valve damage, check the exhaust valve is damaged, replace the bad exhaust valve.

Second, the rotary vane vacuum pump noise: pump noise here.

1, Knocking cylinder, the pump issued an irregular sound operation, like a metal beat metal sound. This is the voice of the vane that hits the pump body. This situation is mainly caused by the spring break between mating rotors or the contraction or ejection failure. The solution: Open the pump to check whether the rotary spring is damaged. Replace the spring.

2, exhaust valve noise, the main exhaust valve is damaged. Solution: a good replacement of the exhaust valve.

Third, the rotary vane vacuum pump vacuum: refers to the current measured vane vacuum pump vacuum than the factory index or the previous use of low. There are many reasons for the low vacuum level. Here are a few key facts.

1, the added rotary vane vacuum pump oil grades and the original is not the same. Different vacuum pump oil grades, as different grades of saturated vapor pressure within the oil is not the same, so the result is not the same. Solution: According to the product model specifications to replace the correct new vacuum pump oil.

2, but also due to vacuum pump oil caused by the low vacuum situation, that is, vacuum pump oil emulsification discoloration, or too dirty. The solution: Put all the vacuum pump oil pump clean, replace the same type of vacuum pump oil, and to be pumping gas in the water vapor and impurities do not enter the pump.

3, was pumping gas temperature may be too high. The solution: to reduce the temperature of the gas being drawn, or you can add a corresponding heat exchanger.

4, the pump is blocked or poor road, the pump cavity does not maintain a certain amount of oil. Solution: check whether the smooth flow of oil, plus the same type of vacuum pump oil.

5, with the gap to change. This is the long-term exhaust gas containing dust, etc., resulting in rotary vane vacuum pump and the gap between the stator wear increases. Solution: check the gap is too large, replace the new parts.

【Print】 【Close】

  According to the temperature change of the working environment, the thermostat will have some physical deformations within the switch, resulting in some special effect. The thermostat will cause a series of automatic control element of turning off and turning on, which is also called temperature control switch, temperature controller, temperature protector, referred to the thermostat. Through thermal protector, the temperature protector will send the temperature to the thermostat and the thermostat will send the commands to control the operation of the device in order to achieve the desired temperature and the energy saving effect. Now the thermostat are widely used in thermostat application appliances, motors, cooling, heating and many other products.

  The working principle of the thermostat is automatically sampling, real-time monitoring through the temperature sensor. The control circuit starts when the ambient temperature is higher than the control set point. If the temperature still rise, when the temperature rose to over-limit alarm set point, the thermostat start limit alarm function. When temperature control can not be effectively controlled in order to prevent destruction of the device also can be stopped by the trip device features continue to run.

  According to the function of the working,the thermostat have different types:Intelligent Thermostat,Remote Controlled Thermostat,Programmable Thermostat,Wireless Room Thermostat.



Room Thermostat

Room Thermostat,Auto Thermostat,Intelligent Thermostat,Remote Controlled Thermostat,Programmable Thermostat,Wireless Room Thermostat

Berggruen Technology Co.,Ltd , http://www.meter-technology.com